Inicio » Producto » Fuente de alimentación médica de estructura abierta » Serie de marco abierto con doble salida de 450 vatios
| Modelo | Voltaje de salida | O/P Current(Convection) | O/P Current(With Fan) | O/P para ventilador | Watt(Convection) | Watt(With Fan) |
|---|---|---|---|---|---|---|
| ATM450S-F120 | 12,0 V | 16,0 A | 25,0 A | 12 V/0,6 A | 260W | 462W |
| ATM450S-F120 | 5,00 V | 1.0A | 1.0A | 12 V/0,6 A | 260W | 462W |
| ATM450S-F150 | 15,0 V | 16,0 A | 25,0 A | 12 V/0,6 A | 260W | 462W |
| ATM450S-F150 | 5,00 V | 1.0A | 1.0A | 12 V/0,6 A | 260W | 462W |
| ATM450S-F190 | 19,0 V | 16,0 A | 25,0 A | 12 V/0,6 A | 260W | 462W |
| ATM450S-F190 | 5,00 V | 1.0A | 1.0A | 12 V/0,6 A | 260W | 462W |
| ATM450S-F240 | 24,0 V | 16,0 A | 25,0 A | 12 V/0,6 A | 260W | 463W |
| ATM450S-F240 | 5,00 V | 1.0A | 1.0A | 12 V/0,6 A | 260W | 463W |
| ATM450S-F360 | 36,0 V | 16,0 A | 25,0 A | 12 V/0,6 A | 260W | 462W |
| ATM450S-F360 | 5,00 V | 1.0A | 1.0A | 12 V/0,6 A | 260W | 462W |
| ATM450S-F480 | 48,0 V | 16,0 A | 25,0 A | 12 V/0,6 A | 260W | 463W |
| ATM450S-F480 | 5,00 V | 1.0A | 1.0A | 12 V/0,6 A | 260W | 463W |
| ATM450S-F540 | 54.0V | 16,0 A | 25,0 A | 12 V/0,6 A | 260W | 460W |
| ATM450S-F540 | 5,00 V | 1.0A | 1.0A | 12 V/0,6 A | 260W | 460W |
| ATM450S-F560 | 56,0 V | 16,0 A | 25,0 A | 12 V/0,6 A | 260W | 460W |
| ATM450S-F560 | 5,00 V | 1.0A | 1.0A | 12 V/0,6 A | 260W | 460W |
| Entrada | |
|---|---|
| Voltaje | 100-240 VCA |
| Frecuencia de línea | 50-60 Hz |
| Actual | 5,3 A - 2,2 A |
| Protección | Fusible interno de corriente primaria |
| salida | |
|---|---|
| Regulación de carga | ± 5% (típico) |
| Ondulación y ruido | ≤ 11 TP3T Vp-p a plena carga |
| Respuesta transitoria (dinámica) | 0,5 ms con cambio de carga 50% |
| Tiempo de arranque | < 3000 ms |
| Tiempo de espera | ≥ 10 ms a plena carga |
| Tiempo de subida | < 50 ms |
| Protección | Protección contra cortocircuitos/Protección contra sobretensión/Protección contra sobrecorriente |
| Eléctrico | |
|---|---|
| Resistencia dieléctrica | 4000 VCA Primario - Secundario |
| Corriente de contacto (paciente) Corriente de fuga | < 100 μA |
| Earth Leakage Curent | < 5 mA |
| Eficiencia | Cumple con el nivel VI del Departamento de Energía, la fase 2 de ErP y el nivel 2 de CoC. |
| MTBF | 300 000 horas calculadas a 25 °C, según Telcordia SR-332. |
| Medioambiental | |
|---|---|
| Temperatura de funcionamiento | -20 a +40 °C, |
| Temperatura de almacenamiento | -20 a + 80 °C, |
| Altitud operativa | 5000 m |
| Humedad relativa | Funcionamiento: 20 a 80% RH Almacenamiento: 10 a 90% RH |
| Refrigeración | 250W with Forced Cooling 450W with Forced Cooling |
| Mecánico | |
|---|---|
| Dimension | Open Frame: L 127 x W 76 x H 34 (mm) 5'' x 3'' x 1.3'' |
| Dimension | Enclosed Type(Top Fan): L 127 x W 87 x H 50 (mm) 5'' x 3.4'' x 2.0'' |
| Dimension | Enclosed Type(Side Fan): L 148 x W 80 x H 41 (mm) 5.9'' x 3.2'' x 1.6'' |
| Peso | Open Frame: 390g ( Ref. ) Enclosed Type ( Top Fan ) : 570g ( Ref. ) Enclosed Type ( Side Fan ) : 660g ( Ref. ) |
Dilithink es un fabricante OEM/ODM especializado en soluciones de alimentación médica seguras y estables.
Nos centramos en adaptadores CA/CC, incluida la tecnología GaN de vanguardia para una potencia ultracompacta y de alta densidad.
Todos los productos cumplen con los más altos estándares de seguridad con aislamiento 2X MOPP y cuentan con certificación internacional (IEC/UL 60601-1). Elija Dilithink para obtener una fuente de alimentación fiable en la atención a pacientes críticos.

A la 1 de la madrugada del 19 de octubre de 2021, Apple celebró un evento para anunciar oficialmente el Macbook...

Mantenemos el desarrollo en la industria de los cargadores y estamos desarrollando el cargador PD30W GaN de 30 W...

Introducción a la tecnología de carga de última generación Dilithink ha presentado oficialmente una solución energética de nueva generación, siendo pionera en...

Un legado de especialización: desde 2001, Dilithink se ha dedicado a la energía de grado médico y militar...