{"id":4731,"date":"2026-02-09T14:44:57","date_gmt":"2026-02-09T06:44:57","guid":{"rendered":"https:\/\/medicaladaptertech.com\/medical-industry-news\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\/"},"modified":"2026-04-24T10:49:16","modified_gmt":"2026-04-24T02:49:16","slug":"type-cf-low-leakage-shielded-transformer-llc-medical-psu","status":"publish","type":"post","link":"https:\/\/medicaladaptertech.com\/nb\/medical-industry-news\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\/","title":{"rendered":"How Shielded Transformers and LLC Topology Improve Medical PSU Leakage Performance"},"content":{"rendered":"<figure class=\"wp-block-image aligncenter size-large\"><img fetchpriority=\"high\" decoding=\"async\" width=\"1408\" height=\"768\" src=\"https:\/\/medicaladaptertech.com\/wp-content\/uploads\/2026\/02\/15921a269f43435c81c6233a51aa4c1f.jpg\" alt=\"How Shielded Transformers and LLC Topology Improve Medical PSU Leakage Performance\" class=\"wp-image-4728\" srcset=\"https:\/\/medicaladaptertech.com\/wp-content\/uploads\/2026\/02\/15921a269f43435c81c6233a51aa4c1f.jpg 1408w, https:\/\/medicaladaptertech.com\/wp-content\/uploads\/2026\/02\/15921a269f43435c81c6233a51aa4c1f-300x164.jpg 300w, https:\/\/medicaladaptertech.com\/wp-content\/uploads\/2026\/02\/15921a269f43435c81c6233a51aa4c1f-1024x559.jpg 1024w, https:\/\/medicaladaptertech.com\/wp-content\/uploads\/2026\/02\/15921a269f43435c81c6233a51aa4c1f-768x419.jpg 768w, https:\/\/medicaladaptertech.com\/wp-content\/uploads\/2026\/02\/15921a269f43435c81c6233a51aa4c1f-18x10.jpg 18w, https:\/\/medicaladaptertech.com\/wp-content\/uploads\/2026\/02\/15921a269f43435c81c6233a51aa4c1f-800x436.jpg 800w\" sizes=\"(max-width: 1408px) 100vw, 1408px\" \/><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">Medical safety standards demand strict limits on patient leakage current for Type CF low leakage power supply designs. Engineers face challenges when designing medical devices like ventilators and anaesthesia machines for hospital environments. The protective earth connection and safety insulation must ensure earth leakage current remains minimal. Conventional Y-capacitor solutions conflict with patient protection requirements. Compliance with IEC 60601-1 standards requires a medical-rated power supply that minimizes enclosure leakage current and maximizes insulation. Achieving reliable patient monitoring in medical applications relies on advanced supply design that prioritizes safety and reduces leakage current at the source.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Viktige punkter<\/h2>\n\n\n\n<ul class=\"wp-block-list\">\n<li><p>Medical power supplies must limit patient leakage current to below 10\u00b5A for safety.<\/p><\/li>\n\n\n\n<li><p>Understanding the leakage current equation helps designers maintain high impedance and low leakage.<\/p><\/li>\n\n\n\n<li><p>Stray capacitance between transformer windings can significantly increase leakage current.<\/p><\/li>\n\n\n\n<li><p>Using a Faraday shield can effectively reduce stray capacitance and improve patient safety.<\/p><\/li>\n\n\n\n<li><p>LLC topology <a href=\"https:\/\/medicaladaptertech.com\/nb\/nyheter-fra-medisinsk-industri\/medical-emi-emc-compliance-usa-top-5-power-supply-tips\/\" target=\"_blank\">minimizes electromagnetic interference<\/a> and reduces the need for large Y-capacitors.<\/p><\/li>\n\n\n\n<li><p>Engineers must balance electromagnetic compatibility with patient protection in medical devices.<\/p><\/li>\n\n\n\n<li><p>Modern designs utilize advanced shielding and LLC topology to achieve reliable patient protection.<\/p><\/li>\n\n\n\n<li><p>Regular testing and compliance with <a href=\"https:\/\/medicaladaptertech.com\/nb\/nyheter-fra-medisinsk-industri\/read-iec-60601-1-test-report-power-supply-safety-compliance\/\" target=\"_blank\">IEC 60601-1-standarder<\/a> ensure safety in medical power supplies.<\/p><\/li>\n<\/ul>\n\n\n\n<h2 class=\"wp-block-heading\">The Physics of Leakage Current in Medical Power Supplies<\/h2>\n\n\n\n<h3 class=\"wp-block-heading\">Leakage Current Equation<\/h3>\n\n\n\n<h4 class=\"wp-block-heading\">Explain <strong>I_leak = V \/ Z_total<\/strong>. Emphasize that at 264VAC, impedance must be extremely high.<\/h4>\n\n\n\n<p class=\"wp-block-paragraph\">Medical power supply designers must understand the quantitative relationship between input voltage, impedance, and resulting leakage current. The equation <strong>I_leak = V \/ Z_total<\/strong> describes how leakage current depends on the applied voltage and the total impedance between the patient and the protective earth. In hospital environments, medical devices often operate at voltages up to 264VAC. To comply with <a href=\"https:\/\/medicaladaptertech.com\/nb\/nyheter-fra-medisinsk-industri\/medical-vs-industrial-power-supply-iec-60601-1-patient-safety\/\" target=\"_blank\">IEC 60601-1-standarder<\/a>, the impedance must remain extremely high to keep patient leakage current below the strict threshold. The capacitance and number of Y capacitors directly affect impedance. Larger Y capacitors decrease impedance, which increases leakage current and threatens patient protection.<\/p>\n\n\n\n<blockquote class=\"wp-block-quote is-layout-flow wp-block-quote-is-layout-flow\">\n<p class=\"wp-block-paragraph\">At higher voltages, even small reductions in impedance can cause leakage current to exceed safety limits. Medical supply engineers must carefully select components to maintain high insulation and minimize leakage.<\/p>\n<\/blockquote>\n\n\n\n<p class=\"wp-block-paragraph\">The following table summarizes the factors influencing leakage current in medical power supplies:<\/p>\n\n\n<figure class=\"wp-block-table\">\n<table class=\"has-fixed-layout\">\n<colgroup>\n<col style=\"min-width: 25px;\" \/>\n<col style=\"min-width: 25px;\" \/><\/colgroup>\n<tbody>\n<tr>\n<th colspan=\"1\" rowspan=\"1\">\n<p>Faktor<\/p>\n<\/th>\n<th colspan=\"1\" rowspan=\"1\">\n<p>Beskrivelse<\/p>\n<\/th>\n<\/tr>\n<tr>\n<td colspan=\"1\" rowspan=\"1\">\n<p>Input Voltage<\/p>\n<\/td>\n<td colspan=\"1\" rowspan=\"1\">\n<p>Higher input voltage results in increased leakage current tested.<\/p>\n<\/td>\n<\/tr>\n<tr>\n<td colspan=\"1\" rowspan=\"1\">\n<p>Impedance (Y Capacitors)<\/p>\n<\/td>\n<td colspan=\"1\" rowspan=\"1\">\n<p>The capacitance and number of Y capacitors affect the leakage current; larger values lead to higher leakage.<\/p>\n<\/td>\n<\/tr>\n<tr>\n<td colspan=\"1\" rowspan=\"1\">\n<p>Leakage Current Equation<\/p>\n<\/td>\n<td colspan=\"1\" rowspan=\"1\">\n<p>Leakage current is calculated as I = 2 * \u03c0 * f * C * Vrms, indicating a direct relationship with frequency and capacitance.<\/p>\n<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<\/figure>\n\n\n<h3 class=\"wp-block-heading\">The Role of Stray Capacitance (Cps)<\/h3>\n\n\n\n<h4 class=\"wp-block-heading\">Explain how capacitance forms between Primary and Secondary windings. This &#8220;invisible capacitor&#8221; is the main path for leakage noise.<\/h4>\n\n\n\n<p class=\"wp-block-paragraph\">Stray capacitance, also known as inter-winding capacitance (Cps), forms between the primary and secondary windings of a transformer in a medical power supply. This invisible capacitor creates a path for common mode noise and leakage current to flow from the supply\u2019s primary side to the patient-connected secondary side. The presence of stray capacitance <a href=\"https:\/\/pmc.ncbi.nlm.nih.gov\/articles\/PMC11930975\/\" target=\"_blank\" rel=\"nofollow\">reduces the isolation level between circuits<\/a>, which is crucial for patient safety and protection. Transformer design must address this issue because excessive stray capacitance can impact resonant converters, decrease output power, and compromise supply efficiency. At the end of the charging period, the effect of stray capacitance becomes more pronounced, requiring careful engineering to mitigate its impact.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Medical supply designers use shielding techniques, such as a Faraday shield, to reduce coupling capacitance by approximately 20-30dB. This approach shunts noise back to the primary side, preventing it from crossing the isolation barrier and reaching the patient. Maintaining high insulation and minimizing stray capacitance ensures that the power supply meets strict leakage current requirements for medical devices in hospital settings. The supply must deliver reliable patient protection while complying with IEC 60601-1 standards.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">The Y-Capacitor Dilemma: EMC vs. Patient Safety<\/h2>\n\n\n\n<h3 class=\"wp-block-heading\">Why We Need Y-Caps for EMC<\/h3>\n\n\n\n<h4 class=\"wp-block-heading\">Y-caps shunt Common Mode noise to ground. Without them, passing CISPR 11 Class B is difficult.<\/h4>\n\n\n\n<p class=\"wp-block-paragraph\">Engineers rely on Y-capacitors to suppress common mode noise in medical power supply circuits. These capacitors provide a low-impedance path for high-frequency noise, directing it safely to ground and away from sensitive patient-connected equipment. The use of Y-caps is essential for meeting electromagnetic compatibility requirements, especially in hospital environments where multiple medical devices operate simultaneously. Regulatory standards demand strict control of electromagnetic interference to prevent disruption of patient monitoring and other critical functions.<\/p>\n\n\n\n<blockquote class=\"wp-block-quote is-layout-flow wp-block-quote-is-layout-flow\">\n<p class=\"wp-block-paragraph\"><a href=\"https:\/\/www.emc-directory.com\/community\/what-is-cispr-11\" target=\"_blank\" rel=\"nofollow\">CISPR 11<\/a> Class B sets the benchmark for EMC performance in medical power supply systems. Devices must meet these standards to operate safely in residential and hospital settings.<\/p>\n<\/blockquote>\n\n\n\n<ul class=\"wp-block-list\">\n<li><p>CISPR 11 is an EMC standard for ISM equipment operating up to 400 GHz, covering medical devices and Wi-Fi systems.<\/p><\/li>\n\n\n\n<li><p>Class B devices are suitable for residential environments, which is crucial for medical power supplies used in home healthcare.<\/p><\/li>\n\n\n\n<li><p>Y-capacitors mitigate common mode noise while ensuring compliance with leakage current requirements in Class II medical devices.<\/p><\/li>\n<\/ul>\n\n\n\n<p class=\"wp-block-paragraph\">Medical power supply designers must <a target=\"_blank\" href=\"https:\/\/medicaladaptertech.com\/nb\/nyheter-fra-medisinsk-industri\/low-leakage-current-power-supply-medical-adapter-design\/\">balance the need for EMC<\/a> compliance with patient protection. The supply must suppress common mode noise without compromising insulation or increasing leakage current.<\/p>\n\n\n\n<h3 class=\"wp-block-heading\">Why Type CF Limits Y-Caps<\/h3>\n\n\n\n<h4 class=\"wp-block-heading\">Calculating the limit: For &lt;10\u00b5A leakage, total Y-capacitance must be &lt;100pF (virtually zero). This forces the need for &#8220;Low Noise at Source&#8221; designs.<\/h4>\n\n\n\n<p class=\"wp-block-paragraph\">Patient safety takes priority in Type CF medical devices. IEC 60601-1 standards require that leakage current remain below 10\u00b5A to protect patients from electrical hazards. The insulation barrier and protective earth must work together to minimize leakage. Engineers calculate the maximum allowable Y-capacitance to ensure compliance with these strict requirements.<\/p>\n\n\n\n<ul class=\"wp-block-list\">\n<li><p>The maximum allowable Y-capacitance for Type CF medical devices is approximately <a href=\"https:\/\/www.specap.com\/resources\/guides\/medical-device-capacitors\" target=\"_blank\" rel=\"nofollow\">100 pF<\/a>.<\/p><\/li>\n\n\n\n<li><p>This limit ensures that leakage current remains below 10 \u00b5A.<\/p><\/li>\n\n\n\n<li><p>The specified limit includes PCB parasitics and transformer inter-winding capacitance.<\/p><\/li>\n<\/ul>\n\n\n\n<p class=\"wp-block-paragraph\"><a target=\"_blank\" rel=\"nofollow\" href=\"https:\/\/www.medicaldesignbriefs.com\/component\/content\/article\/35510-medical-power-supplies-and-electromagnetic-interference\">Limiting Y-capacitance<\/a> affects the ability of medical power supplies to pass EMC compliance tests. Smaller Y-capacitors can meet leakage current requirements but may increase electromagnetic interference, potentially jeopardizing EMC compliance. Designers must implement <a target=\"_blank\" href=\"https:\/\/medicaladaptertech.com\/es\/noticias-del-sector-medico\/high-power-density-medical-power-space-constrained-systems\/\">advanced isolation techniques<\/a> and consider complex filtering to balance patient protection and EMI performance. The supply must deliver reliable power while maintaining high insulation and minimizing leakage. Engineers often migrate to low-noise topologies and shielded transformer designs to achieve these goals.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Solution A: Advanced Shielded Transformer Design<\/h2>\n\n\n\n<h3 class=\"wp-block-heading\">The Faraday Shield Technique<\/h3>\n\n\n\n<h4 class=\"wp-block-heading\">Deep dive: Placing a grounded Copper Foil between Primary and Secondary windings to intercept capacitive current.<\/h4>\n\n\n\n<p class=\"wp-block-paragraph\">Engineers in the medical field often face the challenge of <a href=\"https:\/\/medicaladaptertech.com\/nb\/nyheter-fra-medisinsk-industri\/low-leakage-current-power-supply-medical-adapter-design\/\" target=\"_blank\">minimizing patient leakage<\/a> in power supply systems. The Faraday shield technique addresses this by introducing a grounded copper foil between the primary and secondary windings of the transformer. This shield acts as a physical barrier, intercepting capacitive current that would otherwise flow through the inter-winding capacitance (Cps). By connecting the shield to the protective earth, the design redirects common mode noise and high-frequency dV\/dt transients back to the primary side, away from the patient circuit.<\/p>\n\n\n\n<blockquote class=\"wp-block-quote is-layout-flow wp-block-quote-is-layout-flow\">\n<p class=\"wp-block-paragraph\">The Faraday shield does not carry load current. Instead, it serves as a dedicated path for displacement current, ensuring that noise and leakage do not cross the isolation barrier.<\/p>\n<\/blockquote>\n\n\n\n<p class=\"wp-block-paragraph\">Medical transformers serve as <a target=\"_blank\" rel=\"nofollow\" href=\"https:\/\/www.powersystemsdesign.com\/articles\/leakage-current-testing-requirements-for-medical-grade-transformers\/22\/12174\">isolation devices<\/a>, providing protection for both patients and staff. Safety shields, such as the Faraday shield, help maintain leakage current within the <a target=\"_blank\" href=\"https:\/\/medicaladaptertech.com\/nb\/nyheter-fra-medisinsk-industri\/leakage-current-requirements-medical-power-adapters-sweden\/\">strict limits set by IEC 60601-1<\/a>. In hospital environments, this technique proves essential for patient safety, especially in Type CF medical devices where the allowable leakage current is less than 10\u03bcA.<\/p>\n\n\n\n<h3 class=\"wp-block-heading\">Reducing Coupling Capacitance<\/h3>\n\n\n\n<h4 class=\"wp-block-heading\">How shielding reduces Cps effectively, shunting noise back to the primary side instead of letting it cross the isolation barrier.<\/h4>\n\n\n\n<p class=\"wp-block-paragraph\">The effectiveness of the Faraday shield lies in its ability to reduce the inter-winding capacitance (Cps) between the transformer windings. Lower Cps means less capacitive coupling, which directly translates to reduced leakage current in the power supply. Shielding sensitive components in the transformer design limits the transfer of electromagnetic interference and leakage paths.<\/p>\n\n\n\n<ul class=\"wp-block-list\">\n<li><p>Implementing advanced isolation techniques, such as transformer-based shielding, can significantly limit leakage current transfer.<\/p><\/li>\n\n\n\n<li><p>Shielded transformer designs have demonstrated the ability to reduce patient leakage current from levels above 10\u03bcA to well below the threshold, with some projects achieving reductions from 11\u03bcA to 6\u03bcA.<\/p><\/li>\n\n\n\n<li><p>Adding an extra DC-DC converter further enhances isolation, making it possible for the power supply to achieve patient leakage currents below 10\u03bcA.<\/p><\/li>\n<\/ul>\n\n\n\n<p class=\"wp-block-paragraph\">The following table summarizes the impact of advanced shielding on leakage current in medical power supply systems:<\/p>\n\n\n\n<figure class=\"wp-block-table\">\n<table class=\"has-fixed-layout\">\n<colgroup><col style=\"min-width: 25px;\"\/><col style=\"min-width: 25px;\"\/><col style=\"min-width: 25px;\"\/><\/colgroup><tbody><tr><th colspan=\"1\" rowspan=\"1\"><p>Teknikk<\/p><\/th><th colspan=\"1\" rowspan=\"1\"><p>Typical Leakage Current Achieved<\/p><\/th><th colspan=\"1\" rowspan=\"1\"><p>Application in Medical Devices<\/p><\/th><\/tr><tr><td colspan=\"1\" rowspan=\"1\"><p>Standard Transformer<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>&gt;10\u03bcA<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>Limited use in Type CF<\/p><\/td><\/tr><tr><td colspan=\"1\" rowspan=\"1\"><p>Shielded Transformer<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>&lt;10\u03bcA<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>Suitable for Type CF<\/p><\/td><\/tr><tr><td colspan=\"1\" rowspan=\"1\"><p>Shield + Extra DC-DC<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>6\u03bcA<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>High safety margin<\/p><\/td><\/tr><\/tbody>\n<\/table>\n<\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">By reducing coupling capacitance and shunting noise back to the primary side, shielded transformer designs ensure that the power supply meets the stringent requirements of IEC 60601-1. This approach provides reliable patient protection and maintains high insulation, supporting the safe operation of medical devices in hospital settings.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Solution B: Migrating to LLC Resonant Topology<\/h2>\n\n\n\n<h3 class=\"wp-block-heading\">Limitations of Flyback (Hard Switching)<\/h3>\n\n\n\n<h4 class=\"wp-block-heading\">Flyback converters generate high dV\/dt noise spikes, requiring large Y-caps to fix. Not ideal for Type CF.<\/h4>\n\n\n\n<p class=\"wp-block-paragraph\">Flyback topologies have served as a standard in medical power supply design for decades. However, they present significant challenges for patient protection in Type CF applications. Hard-switching flyback converters generate high <a href=\"https:\/\/medicaladaptertech.com\/nb\/nyheter-fra-medisinsk-industri\/low-leakage-current-power-supply-medical-adapter-design\/\" target=\"_blank\">dV\/dt noise<\/a> spikes during switching transitions. These spikes create substantial common mode noise, which can couple through inter-winding capacitance (Cps) in the transformer. The result is an increase in leakage current that can bypass the insulation barrier and reach the patient side.<\/p>\n\n\n\n<ul class=\"wp-block-list\">\n<li><p>Hard-switching flyback converters produce significant dV\/dt noise, complicating compliance with leakage current standards.<\/p><\/li>\n\n\n\n<li><p>Parasitic capacitance, especially from the transformer, forms a capacitive path for AC current. This path can bypass galvanic isolation and increase patient leakage.<\/p><\/li>\n\n\n\n<li><p>The equation $I_{leakage} = 2pi f C V$ shows that leakage current depends on the capacitance between primary and secondary sides, the frequency, and the applied voltage.<\/p><\/li>\n<\/ul>\n\n\n\n<p class=\"wp-block-paragraph\">To suppress this noise and meet electromagnetic compatibility requirements, engineers often add large Y-capacitors between primary and secondary circuits. However, in medical devices for hospital environments, especially those requiring Type CF protection, the use of large Y-caps is not feasible. The IEC 60601-1 standard limits total Y-capacitance to ensure patient leakage current remains below 10\u00b5A. This restriction forces designers to seek alternative solutions that do not compromise patient safety or insulation.<\/p>\n\n\n\n<h3 class=\"wp-block-heading\">Advantages of LLC (Soft Switching)<\/h3>\n\n\n\n<h4 class=\"wp-block-heading\">LLC topology uses Zero Voltage Switching (ZVS). The sine-wave current creates naturally low EMI, reducing the need for Y-caps in the first place.<\/h4>\n\n\n\n<p class=\"wp-block-paragraph\">LLC resonant topology offers a transformative approach for medical power supply design. This topology enables soft switching, specifically Zero Voltage Switching (ZVS), which allows power devices to switch when voltage across them is nearly zero. As a result, the supply generates minimal dV\/dt noise and significantly reduces common mode noise. The sine-wave current profile of the LLC converter further minimizes high-frequency harmonics, leading to lower electromagnetic interference (EMI).<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">LLC resonant converters operate with reduced current and voltage stress on switching devices. This characteristic extends the lifespan of the power supply and enhances reliability in medical applications. The reduction in EMI means that designers can use much smaller Y-capacitors, or in some cases, eliminate them entirely. This directly supports compliance with the strict leakage current limits required for patient protection in Type CF medical devices.<\/p>\n\n\n\n<blockquote class=\"wp-block-quote is-layout-flow wp-block-quote-is-layout-flow\">\n<p class=\"wp-block-paragraph\">LLC topology <a href=\"https:\/\/www.sciencedirect.com\/science\/article\/pii\/S2772671124003504\" target=\"_blank\" rel=\"nofollow\">improves power quality<\/a> by reducing harmonics, fluctuations, and output voltage distortions. This results in a cleaner and more stable supply, which is essential for sensitive patient monitoring equipment in hospital settings.<\/p>\n<\/blockquote>\n\n\n\n<p class=\"wp-block-paragraph\"><a target=\"_blank\" rel=\"nofollow\" href=\"https:\/\/www.elevify.com\/en-ca\/courses\/engineering-construction-and-technology\/electronics\/switchmode-power-supply-course-11413\">The following table compares<\/a> key aspects of flyback and LLC topologies in medical power supply applications:<\/p>\n\n\n\n<figure class=\"wp-block-table\">\n<table class=\"has-fixed-layout\">\n<colgroup><col style=\"min-width: 25px;\"\/><col style=\"min-width: 25px;\"\/><col style=\"min-width: 25px;\"\/><\/colgroup><tbody><tr><th colspan=\"1\" rowspan=\"1\"><p>Aspekt<\/p><\/th><th colspan=\"1\" rowspan=\"1\"><p>Flyback Topology<\/p><\/th><th colspan=\"1\" rowspan=\"1\"><p>LLC Topology<\/p><\/th><\/tr><tr><td colspan=\"1\" rowspan=\"1\"><p>EMI<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>H\u00f8y<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>Lav<\/p><\/td><\/tr><tr><td colspan=\"1\" rowspan=\"1\"><p>dV\/dt Noise<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>H\u00f8y<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>Lav<\/p><\/td><\/tr><tr><td colspan=\"1\" rowspan=\"1\"><p>Y-Capacitor Size<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>Large<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>Small\/None<\/p><\/td><\/tr><tr><td colspan=\"1\" rowspan=\"1\"><p>Suitability for Type CF<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>Limited<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>Utmerket<\/p><\/td><\/tr><\/tbody>\n<\/table>\n<\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">By <a target=\"_blank\" href=\"https:\/\/medicaladaptertech.com\/pt\/noticias\/\">migrating to LLC resonant topology<\/a>, engineers achieve high insulation, maintain protective earth integrity, and ensure that the supply meets IEC 60601-1 standards. This approach delivers robust patient protection, reliable power, and low leakage current for advanced medical devices.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Comparing Traditional vs. Modern Medical PSU Designs<\/h2>\n\n\n\n<figure class=\"wp-block-image aligncenter size-large\"><img decoding=\"async\" width=\"1200\" height=\"675\" src=\"https:\/\/medicaladaptertech.com\/wp-content\/uploads\/2026\/02\/effab954bfc744aea7efd02fe4665eda.webp\" alt=\"Comparing Traditional vs. Modern Medical PSU Designs\" class=\"wp-image-4729\" srcset=\"https:\/\/medicaladaptertech.com\/wp-content\/uploads\/2026\/02\/effab954bfc744aea7efd02fe4665eda.webp 1200w, https:\/\/medicaladaptertech.com\/wp-content\/uploads\/2026\/02\/effab954bfc744aea7efd02fe4665eda-300x169.webp 300w, https:\/\/medicaladaptertech.com\/wp-content\/uploads\/2026\/02\/effab954bfc744aea7efd02fe4665eda-1024x576.webp 1024w, https:\/\/medicaladaptertech.com\/wp-content\/uploads\/2026\/02\/effab954bfc744aea7efd02fe4665eda-768x432.webp 768w, https:\/\/medicaladaptertech.com\/wp-content\/uploads\/2026\/02\/effab954bfc744aea7efd02fe4665eda-18x10.webp 18w, https:\/\/medicaladaptertech.com\/wp-content\/uploads\/2026\/02\/effab954bfc744aea7efd02fe4665eda-800x450.webp 800w\" sizes=\"(max-width: 1200px) 100vw, 1200px\" \/><figcaption class=\"wp-element-caption\">Bildekilde: <a target=\"_blank\" rel=\"nofollow\" href=\"https:\/\/unsplash.com\">unsplash<\/a><\/figcaption><\/figure>\n\n\n\n<h3 class=\"wp-block-heading\">The Old Way (Flyback + Large Y-Caps)<\/h3>\n\n\n\n<h4 class=\"wp-block-heading\">Good for Type B\/BF, but fails Type CF. High stress on isolation.<\/h4>\n\n\n\n<p class=\"wp-block-paragraph\">Traditional medical power supply designs relied on flyback converters paired with large Y-capacitors. Engineers used this approach to suppress common mode noise and meet electromagnetic compatibility requirements in hospital environments. Flyback topology produced high dV\/dt noise spikes, which increased the need for Y-capacitors to shunt noise to ground. This method worked well for Type B and Type BF medical devices, where patient leakage current limits were less stringent. Type B required <a target=\"_blank\" rel=\"nofollow\" href=\"https:\/\/24x7mag.com\/standards\/regulations\/no-comparison-alternatives-in-medical-device-insulation-testing\/\">isolation at 1500 Vac<\/a>, and Type BF at 3000 Vac. These levels allowed larger Y-capacitance without exceeding safety thresholds.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">However, this strategy placed significant stress on the insulation barrier. Large Y-capacitors reduced impedance between primary and secondary circuits, increasing leakage current. For Type CF applications, which demand patient leakage below 10\u00b5A and insulation at 4000 Vac, traditional designs often failed to comply with IEC 60601-1 standards. The high inter-winding capacitance (Cps) and resistive leakage paths made it difficult to achieve reliable patient protection.<\/p>\n\n\n\n<blockquote class=\"wp-block-quote is-layout-flow wp-block-quote-is-layout-flow\">\n<p class=\"wp-block-paragraph\">Engineers observed that traditional designs often produced lower resistive leakage, but switch mode supplies introduced higher capacitive and inductive leakage currents.<\/p>\n<\/blockquote>\n\n\n\n<figure class=\"wp-block-table\">\n<table class=\"has-fixed-layout\">\n<colgroup><col style=\"min-width: 25px;\"\/><col style=\"min-width: 25px;\"\/><col style=\"min-width: 25px;\"\/><\/colgroup><tbody><tr><th colspan=\"1\" rowspan=\"1\"><p>Aspekt<\/p><\/th><th colspan=\"1\" rowspan=\"1\"><p>Traditional Designs<\/p><\/th><th colspan=\"1\" rowspan=\"1\"><p>Modern Designs<\/p><\/th><\/tr><tr><td colspan=\"1\" rowspan=\"1\"><p>Insulation Testing Method<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>DC insulation tests<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>AC supply measurements<\/p><\/td><\/tr><tr><td colspan=\"1\" rowspan=\"1\"><p>Lekkasjestr\u00f8mtype<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>Primarily resistive leakage<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>Capacitive and inductive leakage<\/p><\/td><\/tr><tr><td colspan=\"1\" rowspan=\"1\"><p>Typical Leakage Values<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>Not specified, often lower<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>Potentially high due to switch mode supplies<\/p><\/td><\/tr><\/tbody>\n<\/table>\n<\/figure>\n\n\n\n<h3 class=\"wp-block-heading\">The DILITHINK Way (LLC + Shielding)<\/h3>\n\n\n\n<h4 class=\"wp-block-heading\">Achieving &lt;10\u00b5A leakage with high margin, ensuring safety even under Single Fault Conditions (SFC).<\/h4>\n\n\n\n<p class=\"wp-block-paragraph\">Modern medical power supply designs use LLC resonant topology and shielded transformers to address the limitations of traditional methods. The LLC approach enables soft switching, which minimizes dV\/dt and reduces common mode noise at the source. Shielded transformers, equipped with a Faraday shield, intercept capacitive currents and lower inter-winding capacitance (Cps) by approximately 20-30dB. This combination allows engineers to minimize Y-capacitance and maintain high insulation, ensuring patient protection.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Medical devices designed with LLC topology and advanced shielding consistently achieve patient leakage currents below 10\u00b5A. These supplies maintain compliance with IEC 60601-1, even under single fault conditions. Safety features such as <a target=\"_blank\" href=\"https:\/\/medicaladaptertech.com\/nb\/nyheter-fra-medisinsk-industri\/medical-power-supply-fusing-iec-60601-1-single-fault-safety\/\">medical power supply fusing<\/a>, dual fusing, active over-current protection, and thermal cutoffs provide additional layers of protection. These components ensure that no hazardous situation arises from a single component failure.<\/p>\n\n\n\n<figure class=\"wp-block-table\">\n<table class=\"has-fixed-layout\">\n<colgroup><col style=\"min-width: 25px;\"\/><col style=\"min-width: 25px;\"\/><\/colgroup><tbody><tr><th colspan=\"1\" rowspan=\"1\"><p>Komponent<\/p><\/th><th colspan=\"1\" rowspan=\"1\"><p>Beskrivelse<\/p><\/th><\/tr><tr><td colspan=\"1\" rowspan=\"1\"><p>Medical Power Supply Fusing<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>Ensures that no hazardous situation arises from a single component failure.<\/p><\/td><\/tr><tr><td colspan=\"1\" rowspan=\"1\"><p>Dual Fusing<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>Provides an additional layer of protection against faults.<\/p><\/td><\/tr><tr><td colspan=\"1\" rowspan=\"1\"><p>Active Over-Current Protection<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>Prevents excessive current that could lead to hazardous conditions.<\/p><\/td><\/tr><tr><td colspan=\"1\" rowspan=\"1\"><p>Thermal Cutoffs<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>Automatically disconnects power in case of overheating, preventing potential hazards.<\/p><\/td><\/tr><\/tbody>\n<\/table>\n<\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">Modern medical power supply designs deliver reliable patient protection and maintain high insulation. Engineers measure leakage current using AC supply measurements, which accurately capture capacitive and inductive leakage. The supply achieves robust performance in hospital environments, supporting advanced medical devices and ensuring safety for patients.<\/p>\n\n\n\n<blockquote class=\"wp-block-quote is-layout-flow wp-block-quote-is-layout-flow\">\n<p class=\"wp-block-paragraph\">The DILITHINK approach demonstrates that minimizing leakage at the source, combined with rigorous safety features, enables medical power supplies to meet the strict requirements of Type CF applications.<\/p>\n<\/blockquote>\n\n\n\n<h2 class=\"wp-block-heading\">Verification and Testing: IEC 60601-1 Compliance<\/h2>\n\n\n\n<h3 class=\"wp-block-heading\">The Measuring Device (MD) Circuit<\/h3>\n\n\n\n<h4 class=\"wp-block-heading\">Explaining the RC network (1k\u03a9 impedance) used in standard tests (IEC 60601-1 Fig. 12).<\/h4>\n\n\n\n<p class=\"wp-block-paragraph\">Medical power supply verification relies on precise measurement of leakage current. IEC 60601-1 defines the Measuring Device (MD) circuit in Figure 12, which simulates the impedance of the human body during testing. The MD circuit consists of a resistor-capacitor (RC) network, typically with a 1k\u03a9 resistor in series with a 0.15\u00b5F capacitor. This configuration creates a total impedance that mimics the electrical characteristics of a patient, ensuring accurate assessment of leakage current from the supply.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Engineers connect the MD circuit between the patient connection and protective earth. The supply operates at maximum rated voltage, and the MD circuit measures the current that flows through the simulated patient path. This method allows for quantitative evaluation of inter-winding capacitance (Cps), common mode noise, and dV\/dt effects. The RC network ensures that the supply meets the strict leakage current limits required for medical devices in hospital environments.<\/p>\n\n\n\n<blockquote class=\"wp-block-quote is-layout-flow wp-block-quote-is-layout-flow\">\n<p class=\"wp-block-paragraph\">The MD circuit provides a standardized approach for testing, enabling comparison across different power supply designs and ensuring consistent protection for patients.<\/p>\n<\/blockquote>\n\n\n<figure class=\"wp-block-table\">\n<table class=\"has-fixed-layout\">\n<colgroup>\n<col style=\"min-width: 25px;\" \/>\n<col style=\"min-width: 25px;\" \/><\/colgroup>\n<tbody>\n<tr>\n<th colspan=\"1\" rowspan=\"1\">\n<p><a href=\"https:\/\/medicaladaptertech.com\/nb\/nyheter-fra-medisinsk-industri\/read-iec-60601-1-test-report-power-supply-safety-compliance\/\" target=\"_blank\" rel=\"noopener\">Evaluation Criteria<\/a><\/p>\n<\/th>\n<th colspan=\"1\" rowspan=\"1\">\n<p>Beskrivelse<\/p>\n<\/th>\n<\/tr>\n<tr>\n<td colspan=\"1\" rowspan=\"1\">\n<p>Limiting leakage currents<\/p>\n<\/td>\n<td colspan=\"1\" rowspan=\"1\">\n<p>Maintains safe levels for patients and operators<\/p>\n<\/td>\n<\/tr>\n<tr>\n<td colspan=\"1\" rowspan=\"1\">\n<p>Protection against electric shock<\/p>\n<\/td>\n<td colspan=\"1\" rowspan=\"1\">\n<p>Evaluates safety under normal and single fault conditions<\/p>\n<\/td>\n<\/tr>\n<tr>\n<td colspan=\"1\" rowspan=\"1\">\n<p>Insulation of application parts<\/p>\n<\/td>\n<td colspan=\"1\" rowspan=\"1\">\n<p>Focuses on safety-critical components<\/p>\n<\/td>\n<\/tr>\n<tr>\n<td colspan=\"1\" rowspan=\"1\">\n<p>Potential equalization<\/p>\n<\/td>\n<td colspan=\"1\" rowspan=\"1\">\n<p>Prevents hazardous voltage differences<\/p>\n<\/td>\n<\/tr>\n<tr>\n<td colspan=\"1\" rowspan=\"1\">\n<p>Safety in medical electrical systems<\/p>\n<\/td>\n<td colspan=\"1\" rowspan=\"1\">\n<p>Ensures system-level integration meets standards<\/p>\n<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<\/figure>\n\n\n<h3 class=\"wp-block-heading\">Single Fault Condition (SFC) Analysis<\/h3>\n\n\n\n<h4 class=\"wp-block-heading\">Testing leakage when the Protective Earth is open (Open Earth). This is the critical &#8220;Pass\/Fail&#8221; moment for German safety officers.<\/h4>\n\n\n\n<p class=\"wp-block-paragraph\">Medical power supply testing extends beyond normal operation. IEC 60601-1 requires evaluation under <a target=\"_blank\" href=\"https:\/\/medicaladaptertech.com\/nb\/nyheter-fra-medisinsk-industri\/medical-power-supply-fusing-iec-60601-1-single-fault-safety\/\">Single Fault Condition<\/a> (SFC), where engineers intentionally disconnect the protective earth. This scenario simulates a fault in the hospital environment, challenging the supply to maintain patient protection even when insulation is compromised.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">SFC analysis identifies potential failure modes and verifies that the supply limits leakage current to safe values. The standard mandates that medical devices remain safe under SFC, with patient leakage current not exceeding 50\u00b5A for Type CF applications. German safety officers often focus on this test, as it represents the most stringent requirement for patient protection.<\/p>\n\n\n\n<figure class=\"wp-block-table\">\n<table class=\"has-fixed-layout\">\n<colgroup><col style=\"min-width: 25px;\"\/><col style=\"min-width: 25px;\"\/><col style=\"min-width: 25px;\"\/><\/colgroup><tbody><tr><th colspan=\"1\" rowspan=\"1\"><p>Equipment Class<\/p><\/th><th colspan=\"1\" rowspan=\"1\"><p>Tilstand<\/p><\/th><th colspan=\"1\" rowspan=\"1\"><p>Lekkasjestr\u00f8mbegrensning<\/p><\/th><\/tr><tr><td colspan=\"1\" rowspan=\"1\"><p>Klasse I<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>Normal tilstand (NC)<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>0.5 mA<\/p><\/td><\/tr><tr><td colspan=\"1\" rowspan=\"1\"><p>Klasse I<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>Enkeltfeiltilstand (SFC)<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>1 mA<\/p><\/td><\/tr><tr><td colspan=\"1\" rowspan=\"1\"><p>Class II<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>Normal tilstand (NC)<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>0.1 mA<\/p><\/td><\/tr><tr><td colspan=\"1\" rowspan=\"1\"><p>Class II<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>Enkeltfeiltilstand (SFC)<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>0.5 mA<\/p><\/td><\/tr><tr><td colspan=\"1\" rowspan=\"1\"><p>Type CF<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>Normal tilstand (NC)<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>0.01 mA<\/p><\/td><\/tr><tr><td colspan=\"1\" rowspan=\"1\"><p>Type CF<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>Enkeltfeiltilstand (SFC)<\/p><\/td><td colspan=\"1\" rowspan=\"1\"><p>0.05 mA<\/p><\/td><\/tr><\/tbody>\n<\/table>\n<\/figure>\n\n\n\n<figure class=\"wp-block-image aligncenter size-large\"><img decoding=\"async\" width=\"1024\" height=\"768\" src=\"https:\/\/medicaladaptertech.com\/wp-content\/uploads\/2026\/02\/chart_1770619071343120370.webp\" alt=\"Dilithink medical power adapters for customized OEM solutions and advanced medical device applicatio.\" class=\"wp-image-4730\" srcset=\"https:\/\/medicaladaptertech.com\/wp-content\/uploads\/2026\/02\/chart_1770619071343120370.webp 1024w, https:\/\/medicaladaptertech.com\/wp-content\/uploads\/2026\/02\/chart_1770619071343120370-300x225.webp 300w, https:\/\/medicaladaptertech.com\/wp-content\/uploads\/2026\/02\/chart_1770619071343120370-768x576.webp 768w, https:\/\/medicaladaptertech.com\/wp-content\/uploads\/2026\/02\/chart_1770619071343120370-16x12.webp 16w, https:\/\/medicaladaptertech.com\/wp-content\/uploads\/2026\/02\/chart_1770619071343120370-800x600.webp 800w\" sizes=\"(max-width: 1024px) 100vw, 1024px\" \/><figcaption class=\"wp-element-caption\">Dilithink specializes in high-quality medical adapters, ensuring reliable power solutions for healthcare equipment.<\/figcaption><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">Medical power supply designs must demonstrate robust insulation and low inter-winding capacitance. The supply must suppress common mode noise and minimize dV\/dt effects to ensure compliance. Engineers use the MD circuit and SFC analysis to verify that the supply delivers reliable protection for patients and operators in hospital settings.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Designers recognize that low leakage in medical power supply systems must originate at the source. Shielded transformer construction, LLC topology, and minimal Y-capacitance form the foundation for reliable protection in hospital environments. Engineers follow a <a target=\"_blank\" href=\"https:\/\/medicaladaptertech.com\/nb\/dilithink-nyheter\/power-supply-inspection-checklist-safety-compliance-guide\/\">checklist to ensure medical devices meet leakage current requirements<\/a>:<\/p>\n\n\n\n<ol class=\"wp-block-list\">\n<li><p>Perform dielectric strength and earth bond tests.<\/p><\/li>\n\n\n\n<li><p>Execute leakage current measurements.<\/p><\/li>\n\n\n\n<li><p>Assess overload and temperature protection.<\/p><\/li>\n\n\n\n<li><p>Compile risk management files and obtain CB Scheme certification.<\/p><\/li>\n<\/ol>\n\n\n\n<p class=\"wp-block-paragraph\">Industry standards such as IEC 60601-1 and IEC 62353 guide ongoing safety. Download DILITHINK\u2019s Type CF Leakage Design Guide for advanced supply solutions.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Ofte stilte sp\u00f8rsm\u00e5l<\/h2>\n\n\n\n<h3 class=\"wp-block-heading\">What is the primary source of leakage current in medical power supplies?<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Inter-winding capacitance (Cps) between transformer windings forms the main path for <a href=\"https:\/\/medicaladaptertech.com\/nb\/nyheter-fra-medisinsk-industri\/leakage-current-requirements-medical-power-adapters-sweden\/\" target=\"_blank\">lekkasjestr\u00f8m<\/a>. This capacitance allows common mode noise to couple from the primary to the secondary circuit, especially at high dV\/dt.<\/p>\n\n\n\n<h3 class=\"wp-block-heading\">How does a Faraday shield reduce leakage current?<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">A Faraday shield, placed between primary and secondary windings, intercepts displacement current. It shunts capacitive noise back to the primary side, reducing Cps by approximately 20\u201330dB.<\/p>\n\n\n\n<h3 class=\"wp-block-heading\">Why does LLC topology improve EMI performance?<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">LLC topology uses soft switching and produces a sinusoidal current waveform. This approach minimizes high-frequency harmonics and dV\/dt, which lowers common mode noise and reduces the need for large Y-capacitors.<\/p>\n\n\n\n<h3 class=\"wp-block-heading\">What is the maximum allowable Y-capacitance for Type CF applications?<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">For Type CF medical devices, total Y-capacitance must remain below 100pF. This limit ensures patient leakage current stays under 10\u00b5A, even at 264VAC input.<\/p>\n\n\n\n<h3 class=\"wp-block-heading\">How do engineers verify compliance with IEC 60601-1 leakage limits?<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Engineers use the Measuring Device (MD) circuit, which simulates human body impedance. They measure leakage current under both normal and single fault conditions to ensure compliance with IEC 60601-1.<\/p>\n\n\n\n<h3 class=\"wp-block-heading\">What happens during Single Fault Condition (SFC) testing?<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">During SFC testing, engineers disconnect protective earth. The power supply must maintain patient leakage below 50\u00b5A for Type CF, demonstrating robust insulation and low inter-winding capacitance.<\/p>\n\n\n\n<h3 class=\"wp-block-heading\">Can shielded transformers alone guarantee compliance with leakage limits?<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Shielded transformers significantly reduce Cps, but compliance also requires low-noise topologies like LLC and minimal Y-capacitance. Engineers must address all sources of leakage current.<\/p>\n\n\n\n<h3 class=\"wp-block-heading\">Why are large Y-capacitors unsuitable for Type CF medical power supplies?<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Large Y-capacitors lower impedance between primary and secondary circuits. This increases leakage current, making it difficult to meet the strict &lt;10\u00b5A requirement for Type CF applications.<\/p>","protected":false},"excerpt":{"rendered":"<p>Shielded transformers and LLC topology enable Type CF low leakage in medical PSUs, meeting IEC 60601-1 standards for patient safety and reliable operation.<\/p>","protected":false},"author":1,"featured_media":4728,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"footnotes":""},"categories":[9],"tags":[199,197,198,196],"class_list":["post-4731","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-medical-industry-news","tag-llc-resonant-converter","tag-low-leakage-current","tag-shielded-transformer","tag-type-cf"],"acf":[],"yoast_head":"<!-- This site is optimized with the Yoast SEO plugin v28.5 - https:\/\/yoast.com\/product\/yoast-seo-wordpress\/ -->\n<title>Type CF Medical PSU: Low Leakage &amp; Shielded LLC Design<\/title>\n<meta name=\"description\" content=\"Discover how shielded transformers and LLC topology achieve ultra-low leakage current for Type CF medical power supplies. Read our 2026 engineering guide.\" \/>\n<meta name=\"robots\" content=\"index, follow, max-snippet:-1, max-image-preview:large, max-video-preview:-1\" \/>\n<link rel=\"canonical\" href=\"https:\/\/medicaladaptertech.com\/nb\/nyheter-fra-medisinsk-industri\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\/\" \/>\n<meta property=\"og:locale\" content=\"nb_NO\" \/>\n<meta property=\"og:type\" content=\"article\" \/>\n<meta property=\"og:title\" content=\"Type CF Medical PSU: Low Leakage &amp; Shielded LLC Design\" \/>\n<meta property=\"og:description\" content=\"Discover how shielded transformers and LLC topology achieve ultra-low leakage current for Type CF medical power supplies. Read our 2026 engineering guide.\" \/>\n<meta property=\"og:url\" content=\"https:\/\/medicaladaptertech.com\/nb\/nyheter-fra-medisinsk-industri\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\/\" \/>\n<meta property=\"og:site_name\" content=\"Dilithink\" \/>\n<meta property=\"article:published_time\" content=\"2026-02-09T06:44:57+00:00\" \/>\n<meta property=\"article:modified_time\" content=\"2026-04-24T02:49:16+00:00\" \/>\n<meta property=\"og:image\" content=\"https:\/\/medicaladaptertech.com\/wp-content\/uploads\/2026\/02\/15921a269f43435c81c6233a51aa4c1f.jpg\" \/>\n\t<meta property=\"og:image:width\" content=\"1408\" \/>\n\t<meta property=\"og:image:height\" content=\"768\" \/>\n\t<meta property=\"og:image:type\" content=\"image\/jpeg\" \/>\n<meta name=\"author\" content=\"Dili\" \/>\n<meta name=\"twitter:card\" content=\"summary_large_image\" \/>\n<meta name=\"twitter:label1\" content=\"Skrevet av\" \/>\n\t<meta name=\"twitter:data1\" content=\"Dili\" \/>\n\t<meta name=\"twitter:label2\" content=\"Ansl. lesetid\" \/>\n\t<meta name=\"twitter:data2\" content=\"15 minutter\" \/>\n<script type=\"application\/ld+json\" class=\"yoast-schema-graph\">{\"@context\":\"https:\\\/\\\/schema.org\",\"@graph\":[{\"@type\":\"Article\",\"@id\":\"https:\\\/\\\/medicaladaptertech.com\\\/medical-industry-news\\\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\\\/#article\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/medicaladaptertech.com\\\/medical-industry-news\\\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\\\/\"},\"author\":{\"name\":\"Dili\",\"@id\":\"https:\\\/\\\/medicaladaptertech.com\\\/#\\\/schema\\\/person\\\/066de28193c805c13a590a9a4af3bc48\"},\"headline\":\"How Shielded Transformers and LLC Topology Improve Medical PSU Leakage Performance\",\"datePublished\":\"2026-02-09T06:44:57+00:00\",\"dateModified\":\"2026-04-24T02:49:16+00:00\",\"mainEntityOfPage\":{\"@id\":\"https:\\\/\\\/medicaladaptertech.com\\\/medical-industry-news\\\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\\\/\"},\"wordCount\":3365,\"commentCount\":0,\"publisher\":{\"@id\":\"https:\\\/\\\/medicaladaptertech.com\\\/#organization\"},\"image\":{\"@id\":\"https:\\\/\\\/medicaladaptertech.com\\\/medical-industry-news\\\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\\\/#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/medicaladaptertech.com\\\/wp-content\\\/uploads\\\/2026\\\/02\\\/15921a269f43435c81c6233a51aa4c1f.jpg\",\"keywords\":[\"LLC Resonant Converter\",\"Low Leakage Current\",\"Shielded Transformer\",\"Type CF\"],\"articleSection\":[\"Medical Industry News\"],\"inLanguage\":\"nb-NO\",\"potentialAction\":[{\"@type\":\"CommentAction\",\"name\":\"Comment\",\"target\":[\"https:\\\/\\\/medicaladaptertech.com\\\/medical-industry-news\\\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\\\/#respond\"]}]},{\"@type\":\"WebPage\",\"@id\":\"https:\\\/\\\/medicaladaptertech.com\\\/medical-industry-news\\\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\\\/\",\"url\":\"https:\\\/\\\/medicaladaptertech.com\\\/medical-industry-news\\\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\\\/\",\"name\":\"Type CF Medical PSU: Low Leakage & Shielded LLC Design\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/medicaladaptertech.com\\\/#website\"},\"primaryImageOfPage\":{\"@id\":\"https:\\\/\\\/medicaladaptertech.com\\\/medical-industry-news\\\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\\\/#primaryimage\"},\"image\":{\"@id\":\"https:\\\/\\\/medicaladaptertech.com\\\/medical-industry-news\\\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\\\/#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/medicaladaptertech.com\\\/wp-content\\\/uploads\\\/2026\\\/02\\\/15921a269f43435c81c6233a51aa4c1f.jpg\",\"datePublished\":\"2026-02-09T06:44:57+00:00\",\"dateModified\":\"2026-04-24T02:49:16+00:00\",\"description\":\"Discover how shielded transformers and LLC topology achieve ultra-low leakage current for Type CF medical power supplies. Read our 2026 engineering guide.\",\"breadcrumb\":{\"@id\":\"https:\\\/\\\/medicaladaptertech.com\\\/medical-industry-news\\\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\\\/#breadcrumb\"},\"inLanguage\":\"nb-NO\",\"potentialAction\":[{\"@type\":\"ReadAction\",\"target\":[\"https:\\\/\\\/medicaladaptertech.com\\\/medical-industry-news\\\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\\\/\"]}]},{\"@type\":\"ImageObject\",\"inLanguage\":\"nb-NO\",\"@id\":\"https:\\\/\\\/medicaladaptertech.com\\\/medical-industry-news\\\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\\\/#primaryimage\",\"url\":\"https:\\\/\\\/medicaladaptertech.com\\\/wp-content\\\/uploads\\\/2026\\\/02\\\/15921a269f43435c81c6233a51aa4c1f.jpg\",\"contentUrl\":\"https:\\\/\\\/medicaladaptertech.com\\\/wp-content\\\/uploads\\\/2026\\\/02\\\/15921a269f43435c81c6233a51aa4c1f.jpg\",\"width\":1408,\"height\":768,\"caption\":\"Dilithink specializes in innovative medical adapters and OEM manufacturing for healthcare technology.\"},{\"@type\":\"BreadcrumbList\",\"@id\":\"https:\\\/\\\/medicaladaptertech.com\\\/medical-industry-news\\\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\\\/#breadcrumb\",\"itemListElement\":[{\"@type\":\"ListItem\",\"position\":1,\"name\":\"Home\",\"item\":\"https:\\\/\\\/medicaladaptertech.com\\\/\"},{\"@type\":\"ListItem\",\"position\":2,\"name\":\"Medical Industry News\",\"item\":\"https:\\\/\\\/medicaladaptertech.com\\\/category\\\/medical-industry-news\\\/\"},{\"@type\":\"ListItem\",\"position\":3,\"name\":\"How Shielded Transformers and LLC Topology Improve Medical PSU Leakage Performance\"}]},{\"@type\":\"WebSite\",\"@id\":\"https:\\\/\\\/medicaladaptertech.com\\\/#website\",\"url\":\"https:\\\/\\\/medicaladaptertech.com\\\/\",\"name\":\"Dilithink\",\"description\":\"Precision Power Supply for Advanced Medical Devices\",\"publisher\":{\"@id\":\"https:\\\/\\\/medicaladaptertech.com\\\/#organization\"},\"alternateName\":\"Dilithink\",\"potentialAction\":[{\"@type\":\"SearchAction\",\"target\":{\"@type\":\"EntryPoint\",\"urlTemplate\":\"https:\\\/\\\/medicaladaptertech.com\\\/?s={search_term_string}\"},\"query-input\":{\"@type\":\"PropertyValueSpecification\",\"valueRequired\":true,\"valueName\":\"search_term_string\"}}],\"inLanguage\":\"nb-NO\"},{\"@type\":\"Organization\",\"@id\":\"https:\\\/\\\/medicaladaptertech.com\\\/#organization\",\"name\":\"Dilithink\",\"alternateName\":\"SHENZHEN DILITHINK ELECTRONICS CO.,LTD.\",\"url\":\"https:\\\/\\\/medicaladaptertech.com\\\/\",\"logo\":{\"@type\":\"ImageObject\",\"inLanguage\":\"nb-NO\",\"@id\":\"https:\\\/\\\/medicaladaptertech.com\\\/#\\\/schema\\\/logo\\\/image\\\/\",\"url\":\"https:\\\/\\\/medicaladaptertech.com\\\/wp-content\\\/uploads\\\/2025\\\/11\\\/DILITHINK-LOGO1_00.png\",\"contentUrl\":\"https:\\\/\\\/medicaladaptertech.com\\\/wp-content\\\/uploads\\\/2025\\\/11\\\/DILITHINK-LOGO1_00.png\",\"width\":709,\"height\":354,\"caption\":\"Dilithink\"},\"image\":{\"@id\":\"https:\\\/\\\/medicaladaptertech.com\\\/#\\\/schema\\\/logo\\\/image\\\/\"}},{\"@type\":\"Person\",\"@id\":\"https:\\\/\\\/medicaladaptertech.com\\\/#\\\/schema\\\/person\\\/066de28193c805c13a590a9a4af3bc48\",\"name\":\"Dili\",\"image\":{\"@type\":\"ImageObject\",\"inLanguage\":\"nb-NO\",\"@id\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/95d5bbd35d36aa5ea46b58a71a6f92aeba931eb539b5ed332061ba01e8b86e37?s=96&d=mm&r=g\",\"url\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/95d5bbd35d36aa5ea46b58a71a6f92aeba931eb539b5ed332061ba01e8b86e37?s=96&d=mm&r=g\",\"contentUrl\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/95d5bbd35d36aa5ea46b58a71a6f92aeba931eb539b5ed332061ba01e8b86e37?s=96&d=mm&r=g\",\"caption\":\"Dili\"},\"sameAs\":[\"https:\\\/\\\/wordpress-1524552-5869595.cloudwaysapps.com\"]}]}<\/script>\n<!-- \/ Yoast SEO plugin. -->","yoast_head_json":{"title":"Type CF Medical PSU: Low Leakage & Shielded LLC Design","description":"Discover how shielded transformers and LLC topology achieve ultra-low leakage current for Type CF medical power supplies. Read our 2026 engineering guide.","robots":{"index":"index","follow":"follow","max-snippet":"max-snippet:-1","max-image-preview":"max-image-preview:large","max-video-preview":"max-video-preview:-1"},"canonical":"https:\/\/medicaladaptertech.com\/nb\/nyheter-fra-medisinsk-industri\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\/","og_locale":"nb_NO","og_type":"article","og_title":"Type CF Medical PSU: Low Leakage & Shielded LLC Design","og_description":"Discover how shielded transformers and LLC topology achieve ultra-low leakage current for Type CF medical power supplies. Read our 2026 engineering guide.","og_url":"https:\/\/medicaladaptertech.com\/nb\/nyheter-fra-medisinsk-industri\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\/","og_site_name":"Dilithink","article_published_time":"2026-02-09T06:44:57+00:00","article_modified_time":"2026-04-24T02:49:16+00:00","og_image":[{"width":1408,"height":768,"url":"https:\/\/medicaladaptertech.com\/wp-content\/uploads\/2026\/02\/15921a269f43435c81c6233a51aa4c1f.jpg","type":"image\/jpeg"}],"author":"Dili","twitter_card":"summary_large_image","twitter_misc":{"Skrevet av":"Dili","Ansl. lesetid":"15 minutter"},"schema":{"@context":"https:\/\/schema.org","@graph":[{"@type":"Article","@id":"https:\/\/medicaladaptertech.com\/medical-industry-news\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\/#article","isPartOf":{"@id":"https:\/\/medicaladaptertech.com\/medical-industry-news\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\/"},"author":{"name":"Dili","@id":"https:\/\/medicaladaptertech.com\/#\/schema\/person\/066de28193c805c13a590a9a4af3bc48"},"headline":"How Shielded Transformers and LLC Topology Improve Medical PSU Leakage Performance","datePublished":"2026-02-09T06:44:57+00:00","dateModified":"2026-04-24T02:49:16+00:00","mainEntityOfPage":{"@id":"https:\/\/medicaladaptertech.com\/medical-industry-news\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\/"},"wordCount":3365,"commentCount":0,"publisher":{"@id":"https:\/\/medicaladaptertech.com\/#organization"},"image":{"@id":"https:\/\/medicaladaptertech.com\/medical-industry-news\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\/#primaryimage"},"thumbnailUrl":"https:\/\/medicaladaptertech.com\/wp-content\/uploads\/2026\/02\/15921a269f43435c81c6233a51aa4c1f.jpg","keywords":["LLC Resonant Converter","Low Leakage Current","Shielded Transformer","Type CF"],"articleSection":["Medical Industry News"],"inLanguage":"nb-NO","potentialAction":[{"@type":"CommentAction","name":"Comment","target":["https:\/\/medicaladaptertech.com\/medical-industry-news\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\/#respond"]}]},{"@type":"WebPage","@id":"https:\/\/medicaladaptertech.com\/medical-industry-news\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\/","url":"https:\/\/medicaladaptertech.com\/medical-industry-news\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\/","name":"Type CF Medical PSU: Low Leakage & Shielded LLC Design","isPartOf":{"@id":"https:\/\/medicaladaptertech.com\/#website"},"primaryImageOfPage":{"@id":"https:\/\/medicaladaptertech.com\/medical-industry-news\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\/#primaryimage"},"image":{"@id":"https:\/\/medicaladaptertech.com\/medical-industry-news\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\/#primaryimage"},"thumbnailUrl":"https:\/\/medicaladaptertech.com\/wp-content\/uploads\/2026\/02\/15921a269f43435c81c6233a51aa4c1f.jpg","datePublished":"2026-02-09T06:44:57+00:00","dateModified":"2026-04-24T02:49:16+00:00","description":"Discover how shielded transformers and LLC topology achieve ultra-low leakage current for Type CF medical power supplies. Read our 2026 engineering guide.","breadcrumb":{"@id":"https:\/\/medicaladaptertech.com\/medical-industry-news\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\/#breadcrumb"},"inLanguage":"nb-NO","potentialAction":[{"@type":"ReadAction","target":["https:\/\/medicaladaptertech.com\/medical-industry-news\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\/"]}]},{"@type":"ImageObject","inLanguage":"nb-NO","@id":"https:\/\/medicaladaptertech.com\/medical-industry-news\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\/#primaryimage","url":"https:\/\/medicaladaptertech.com\/wp-content\/uploads\/2026\/02\/15921a269f43435c81c6233a51aa4c1f.jpg","contentUrl":"https:\/\/medicaladaptertech.com\/wp-content\/uploads\/2026\/02\/15921a269f43435c81c6233a51aa4c1f.jpg","width":1408,"height":768,"caption":"Dilithink specializes in innovative medical adapters and OEM manufacturing for healthcare technology."},{"@type":"BreadcrumbList","@id":"https:\/\/medicaladaptertech.com\/medical-industry-news\/type-cf-low-leakage-shielded-transformer-llc-medical-psu\/#breadcrumb","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https:\/\/medicaladaptertech.com\/"},{"@type":"ListItem","position":2,"name":"Medical Industry News","item":"https:\/\/medicaladaptertech.com\/category\/medical-industry-news\/"},{"@type":"ListItem","position":3,"name":"How Shielded Transformers and LLC Topology Improve Medical PSU Leakage Performance"}]},{"@type":"WebSite","@id":"https:\/\/medicaladaptertech.com\/#website","url":"https:\/\/medicaladaptertech.com\/","name":"Dilithink","description":"Presisjonsstr\u00f8mforsyning for avanserte medisinske enheter","publisher":{"@id":"https:\/\/medicaladaptertech.com\/#organization"},"alternateName":"Dilithink","potentialAction":[{"@type":"SearchAction","target":{"@type":"EntryPoint","urlTemplate":"https:\/\/medicaladaptertech.com\/?s={search_term_string}"},"query-input":{"@type":"PropertyValueSpecification","valueRequired":true,"valueName":"search_term_string"}}],"inLanguage":"nb-NO"},{"@type":"Organization","@id":"https:\/\/medicaladaptertech.com\/#organization","name":"Dilithink","alternateName":"SHENZHEN DILITHINK ELECTRONICS CO.,LTD.","url":"https:\/\/medicaladaptertech.com\/","logo":{"@type":"ImageObject","inLanguage":"nb-NO","@id":"https:\/\/medicaladaptertech.com\/#\/schema\/logo\/image\/","url":"https:\/\/medicaladaptertech.com\/wp-content\/uploads\/2025\/11\/DILITHINK-LOGO1_00.png","contentUrl":"https:\/\/medicaladaptertech.com\/wp-content\/uploads\/2025\/11\/DILITHINK-LOGO1_00.png","width":709,"height":354,"caption":"Dilithink"},"image":{"@id":"https:\/\/medicaladaptertech.com\/#\/schema\/logo\/image\/"}},{"@type":"Person","@id":"https:\/\/medicaladaptertech.com\/#\/schema\/person\/066de28193c805c13a590a9a4af3bc48","name":"Dili","image":{"@type":"ImageObject","inLanguage":"nb-NO","@id":"https:\/\/secure.gravatar.com\/avatar\/95d5bbd35d36aa5ea46b58a71a6f92aeba931eb539b5ed332061ba01e8b86e37?s=96&d=mm&r=g","url":"https:\/\/secure.gravatar.com\/avatar\/95d5bbd35d36aa5ea46b58a71a6f92aeba931eb539b5ed332061ba01e8b86e37?s=96&d=mm&r=g","contentUrl":"https:\/\/secure.gravatar.com\/avatar\/95d5bbd35d36aa5ea46b58a71a6f92aeba931eb539b5ed332061ba01e8b86e37?s=96&d=mm&r=g","caption":"Dili"},"sameAs":["https:\/\/wordpress-1524552-5869595.cloudwaysapps.com"]}]}},"_links":{"self":[{"href":"https:\/\/medicaladaptertech.com\/nb\/wp-json\/wp\/v2\/posts\/4731","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/medicaladaptertech.com\/nb\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/medicaladaptertech.com\/nb\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/medicaladaptertech.com\/nb\/wp-json\/wp\/v2\/users\/1"}],"replies":[{"embeddable":true,"href":"https:\/\/medicaladaptertech.com\/nb\/wp-json\/wp\/v2\/comments?post=4731"}],"version-history":[{"count":0,"href":"https:\/\/medicaladaptertech.com\/nb\/wp-json\/wp\/v2\/posts\/4731\/revisions"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/medicaladaptertech.com\/nb\/wp-json\/wp\/v2\/media\/4728"}],"wp:attachment":[{"href":"https:\/\/medicaladaptertech.com\/nb\/wp-json\/wp\/v2\/media?parent=4731"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/medicaladaptertech.com\/nb\/wp-json\/wp\/v2\/categories?post=4731"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/medicaladaptertech.com\/nb\/wp-json\/wp\/v2\/tags?post=4731"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}