Open Frame Dual Output 450 Watt Series

Model:
ATM450S
Voltage:
5/12V/15V/19V/24V/36V/48V/56V
  • IEC/EN/ANSI/AAMI ES60601-1 (Edition3.2)
  • EMC: IEC60601-1-2 (Edition 4.1)
  • Means of Protection: 2X MOPP
  • Leakage current< 100μA
  • IEC/UL 62368-1
  • 450W With FanF
  • Standby 5V/1A
  • Regulated Output With Low Ripple Noise
  • Modified and Custom Design Available
  • On & Off Remote Control
  • 12V / 0.6A Output for Fan

Product Info

Model:

ModelO/P VoltageO/P Current(Convection)O/P Current(With Fan)O/P for FanWatt(Convection)Watt(With Fan)
ATM450S-F12012.0V16.0A25.0A12V/0.6A260W462W
ATM450S-F1205.00V1.0A1.0A12V/0.6A260W462W
ATM450S-F15015.0V16.0A25.0A12V/0.6A260W462W
ATM450S-F1505.00V1.0A1.0A12V/0.6A260W462W
ATM450S-F19019.0V16.0A25.0A12V/0.6A260W462W
ATM450S-F1905.00V1.0A1.0A12V/0.6A260W462W
ATM450S-F24024.0V16.0A25.0A12V/0.6A260W463W
ATM450S-F2405.00V1.0A1.0A12V/0.6A260W463W
ATM450S-F36036.0V16.0A25.0A12V/0.6A260W462W
ATM450S-F3605.00V1.0A1.0A12V/0.6A260W462W
ATM450S-F48048.0V16.0A25.0A12V/0.6A260W463W
ATM450S-F4805.00V1.0A1.0A12V/0.6A260W463W
ATM450S-F54054.0V16.0A25.0A12V/0.6A260W460W
ATM450S-F5405.00V1.0A1.0A12V/0.6A260W460W
ATM450S-F56056.0V16.0A25.0A12V/0.6A260W460W
ATM450S-F5605.00V1.0A1.0A12V/0.6A260W460W

Input:

Input
Voltage100-240VAC
Line Frequency50-60 Hz
Current5.3A - 2.2A
ProtectionInternal Primary Current Fuse

Output:

output
Load Regulation± 5% ( Typical )
Ripple&Noise≤ 1% Vp-p @ Full Load
Transient(Dynamic) Response0.5mS with 50% Load Change
Start-up Time< 3000mS
Hold-up Time≥ 10mS @ Full Load
Rise Time< 50mS
ProtectionShort Circuit Protection/Over Voltage Protection/Over Current Protection

Electrical:

Electrical
Dielectric Withstand4,000VAC Primary - Secondary
Touch Current(Patient) Leakage Current< 100 μA
Earth Leakage Curent< 5mA
EfficiencyMeet DoE Level VI, ErP Stage 2, CoC Tier 2
MTBF300000 Calculated Hours at 25°C, by Telcordia SR-332

Environmental:

Environmental
Operating Temperature-20 to + 40°C,
Storage Temperature-20 to + 80°C,
Operating Altitude5000M
Relative HumidityOperating: 20 to 80% RH
Storage: 10 to 90% RH
Cooling250W with Forced Cooling
450W with Forced Cooling

Mechanical:

Mechanical
DimensionOpen Frame: L 127 x W 76 x H 34 (mm)
5'' x 3'' x 1.3''
DimensionEnclosed Type(Top Fan): L 127 x W 87 x H 50 (mm)
5'' x 3.4'' x 2.0''
DimensionEnclosed Type(Side Fan): L 148 x W 80 x H 41 (mm)
5.9'' x 3.2'' x 1.6''
WeightOpen Frame: 390g ( Ref. )
Enclosed Type ( Top Fan ) : 570g ( Ref. )
Enclosed Type ( Side Fan ) : 660g ( Ref. )

Dilithink Profile

Dilithink is a dedicated OEM/ODM manufacturer specializing in safe and stable medical power solutions.

We focus on AC/DC adapters, including cutting-edge GaN technology for ultra-compact, high-density power.

All products meet the highest safety standards with 2X MOPP isolation and are globally certified (IEC/UL 60601-1). Choose Dilithink for reliable power in critical patient care.

We have exclusive properties just for you, Leave your details and we'll talk soon.

Lorem ipsum dolor sit amet, consectetur adipiscing elit, sed do eiusmod tempor incid idunt ut labore ellt dolore.